Influence of applied brachial recording forces on pulse wave
velocity and transmission in the brachio-radial arterial segment
M. Darcy Driscoll
J. Malcolm O. Arnold
Gordon E. Marchiori
Marvin H. Sherebrin
Departments of Medicine and Medical Biophysics,
University of Western Ontario, London, Ontario
(Original manuscript submitted 15/3/95; received in revised form
19/6/95; accepted 11/7/95)
Abstract
Arterial stiffness in hypertension and heart failure may increase
afterload on the left ventricle. Pulse wave velocity and transmission
ratio measurements are noninvasive methods to assess arterial
stiffness. Since noninvasive pressure pulse recording requires
sufficient applied force to distort the vessel wall, we hypothesized
that the pulse wave velocity, transmission ratios, and distal pulse
amplitudes and contours may be altered by the recording
technique. Brachial and radial arterial pressure pulses were
recorded simultaneously using a piezoelectric pulse transducer in
14 young, normal, male subjects using 10 brachial artery recording
forces (0.35-3.58N, about 0.36N increments) applied in a random
order and a constant radial force (2.35N). Pulses were Fourier
analyzed. One subject was excluded from analyses because of
improper transducer positioning over the brachial artery. In 8
subjects, no significant changes occurred in any variable over all
brachial recording forces. In the remaining 5 subjects, the
measured variables remained constant until brachial artery
recording forces exceeded 2.42 ± 0.03N. The pulse wave
velocity (p < 0.04), transmission ratios (harmonics 2-5, p <
0.0001), radial pulse amplitude (p < 0.0003), and relative powers
(harmonics 2-5, p < 0.02) then decreased. In these subjects,
brachial artery depths were less than the other subjects (5.9 ±
0.4 vs. 7.7 ± 0.4 mm, p < 0.05). The brachial and radial
artery recording forces normally used during clinical measurements
by 2 investigators were 1.43 ± 0.01N (95% confidence
intervals (CI) = 1.23N, 1.62N) and 1.88 ± 0.11N (95% CI =
1.65N, 2.10N), respectively. Therefore, at forces normally used by
clinical investigators, the pulse wave velocity, harmonic
transmission ratios, and pulse amplitudes and contours obtained at
the brachial and radial artery are not significantly influenced by
forces applied at the brachial artery. However, these variables may
be decreased in subjects with more superficial arteries when higher
recording forces are used.
Clin Invest Med 1995; 18 (6): 435-448
Table of contents: CIM vol. 18, no. 6
Copyright 1996 Canadian Medical Association